Kinematic Analysis of Total Knee Arthroplasty(TKA) Using Orthosensor System: Externally Rotated Femoral Prosthesis Versus Traditional Design Prosthesis

EasyChair Preprint 3777, version 3

Versions: 123history
3 pagesDate: July 24, 2020

Abstract

Purpose The purpose of this study was to elucidate kinematic change according to the implant’s specific femoral rotation by using orthosensor implant with three degrees external rotation of femoral rotation rebuilt and traditional TKA implant without rebuilt of the femoral rotation .
Methods Twenty-eight patients (34 knees) underwent TKA using traditional TKA implant and 16 patients (22 knees) underwent TKA using implant with three degrees external rotation of femoral rotation. Patients were followed up for at least 1 year. Mean age of patients was 71.1 years (range, 60 to 80 years) at the time of surgery. After implantation of femur and tibial components, we applied the orthosensor system, to evaluate femoral rollback of the new artificial joint. Femoral rollback was analyzed using digitized screenshot function of orthosensor system.
Results Overall femoral tracking proportion regardless of implants was significantly higher on the medial compartment compared to that on the lateral compartment (13.3 ± 8.4% vs. 6.3 ± 5.0%, p < 0.001). Regarding femoral tracking according to each compartment, externally rotated femoral prosthesis and traditional prosthesis showed 12.1 ± 8.2% and 14.2 ± 8.6% (p = 0.371) on the medial compartment and 8.0 ± 5.8% and 5.2 ± 4.2% (p = 0.059) on the lateral compartment, respectively.
Conclusion Our study showed reverse femoral roll-back movement with higher tracking distance on the lateral compartment during TKA. externally rotated femoral prosthesis TKA system with femoral component 3-degree rebuilt showed less roll-back difference between medial and lateral compartments compared to traditional TKA system. Fortunately, both TKA systems had excellent short-term clinical outcomes without having significant difference between the two. With longer follow-up and larger cohort, the advantage and effectiveness of femoral component rotation can be elucidated in the future.

Keyphrases: Femoral roll back, Orthosensor, Total knee arthroplasty, kinematic analysis

BibTeX entry
BibTeX does not have the right entry for preprints. This is a hack for producing the correct reference:
@booklet{EasyChair:3777,
  author    = {Seung-Min Na and Gun-Woo Kim and Nam-Hun Lee and Young-Woo Chung and Jongkeun Seon and Eun-Kyoo Song},
  title     = {Kinematic Analysis of Total Knee Arthroplasty(TKA) Using Orthosensor System: Externally Rotated Femoral Prosthesis Versus Traditional Design Prosthesis},
  howpublished = {EasyChair Preprint 3777},
  year      = {EasyChair, 2020}}
Download PDFOpen PDF in browser